ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
128 LIS/HIS:
Failed to
open
connection
The communication
hardware was busy or
the remote system did
not respond.
- Check that the remote
system is running, correctly
configured and responding.
- Check communication
parameters, e.g. baud rate,
parity, IP address, etc., as
defined in Communication
setup.
- Reboot the analyzer.
129 LIS/HIS:
Failed to
close
connection
Messages were queued
when the communica-
tion channel was
closed. Results and
other messages sent by
the analyzer to a
remote system may be
lost.
If the problem persists,
check the communication
hardware. The remote
system may lack buffer
capacity.
131
LIS/HIS:
Failed to
send packet
A communication error
occured while sending
a message. The
message was not sent.
- Check that the remote
system is running and
responding.
- Check communication
hardware including cables.
- Repeat sending.
132 LIS/HIS:
Failed to
receive
packet
An error occured while
receiving a message.
The analyzer was not
able to recognize the
received massage.
- Check that protocol types
are correctly configured on
both the analyzer and the
remote system.
- Contact Radiometer
service representative.
133 LIS/HIS:
Connection
lost
A previously
established LIS/HIS
connection has been
lost.
- Check that the remote
system is running and
responding.
- Check cables.
Continued on next page
11-9