11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
89 Measured
QC value
above the
control
range
Measured parameter
value is above the
control range.
- Verify procedure and
repeat measurement.
- Refer to Quality Control
Systems Reference
Manual.
Crea:
- Mount a new Cleaning
Met II Solution.
- Remembrane both Crea A
and Crea B electrodes
90 Measured
QC value
below
control
range
Measured parameter
value is below the
control range.
- Verify procedure and
repeat measurement.
- Refer to Quality Control
Systems Reference
Manual.
Crea:
- Mount a new Cleaning
Met II Solution.
- Remembrane both Crea A
and Crea B electrodes
93 Value
above the
reportable
range
Parameter value is
above the reportable
range. See chapter 13:
Specifications in the
Operator's Manual
94 Value below
the
reportable
range
Parameter value below
the reportable range.
See chapter 13:
Specifications in the
Operator's Manual.
- Check for and remedy
other errors related to the
result, system messages, or
calibration status.
- Perform QC. If the QC
result is accepted, the
blood sample may be
suspected.
- Repeat measurement on a
new blood sample.
- Perform the Electrode
Troubleshooting procedure.
117 LIS/HIS:
Invalid
connection
configura-
tion
The comunication
configuration or the
protocol definition was
invalid.
Check the communication
parameters specified in
Communications setup.
Continued on next page