ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages
Messages at
manager level
The following messages will be seen at the manager levels. The messages are
listed in numerical order.
Operator actions are listed in order of priority. Perform the first action in the list; if
unsuccessful, try the next action, etc.
No. Message Interpretation Operator action
1 Inconsistent
software
versions.
Please
contact
service.
Inconsistent software
versions for different
modules. May appear
after replacing a
complete module or as
a result of an
incomplete software
upgrade.
Contact Radiometer service
representative.
Removal condition:
Successful software
consistency check.
83 Value above
reference
range
Parameter value is
above the user-defined
reference range.
This is only a message,
not an error.
No action required.
84 Value below
reference
range
Parameter value is
below the user-defined
reference range.
This is only a message,
not an error.
No action required.
85 Value is
below the
critical limit
Parameter value is
below the user-defined
critical limit.
This is only a message,
not an error.
No action required.
86 Value is
above the
critical limit
Parameter value is
above the user-defined
critical limit.
This is only a message,
not an error.
No action required.
Continued on next page
11-7