ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
No. Message Interpretation Operator action
218 Inlet
positioning
error
Inlet probe was not
positioned properly
within required time
frame.
Causes the activity to
abort.
Perform the Inlet
Troubleshooting Procedure
Removal condition:
Successful inlet calibration.
219 Condition-
ing error in
El/Met
electrode
module
Rinse Solution was not
detected in El/Met
electrode module. The
glucose and lactate
results will be reported
as "....." in the next
measurement.
- Check Rinse Solution for
adequate volume, proper
installation or obstructions
in container opening.
Replace as necessary.
- Perform the Fluid
transport troubleshooting
procedure.
Removal condition:
Successful Rinse.
232 Oximetry
calibration
error
Water spectrum
intensity was too low
or too high during
absorbance calculation.
OXI measurements
cannot be performed.
- Perform a 2-point
calibration.
- Contact Radiometer
Service Representative if
the problem persists.
Removal condition:
Successful 2-point
calibration.
234-238,
245-252,
254-258,
269, 314-
316
Oximetry
hardware
error
The Oxi module has a
hardware error.
- Perform a 2-point
calibration.
- Contact Radiometer
Service Representative if
the problem persists.
259, 270,
354-357
Temperature
error
Analyzer temperature
is outside 37.0 ± 0.2
degrees Celsius.
- Ensure the ambient
temperature is between 15
and 32 degrees Celsius.
- If the system has just
performed a cold start, wait
for error to disappear.
- Replace the fan filter if
dirty.
- Shield analyzer from
direct sunlight and other
heat sources.
- Contact Radiometer
service representative.
Messages at
manager level
(continued)
Continued on next page
11-13