11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
290 Warning:
SHb
detected
FSHb detected in the
range of 1-10 %.
No action required.
For information only.
291 SHb too
high
Detected FSHb is
greater than 10 %.
Measurement accuracy
is affected.
Repeat the measurement.
292 Turbidity
too high
Turbidity is greater
than 5 %: too high for
reliable measurements.
- Check pump tubes and
inlet tube for leaks. Repair
as necessary.
- Hyperlipemic sample;
decrease the lipemic
content by, e.g. centrifuge
or extraction.
- Perform the measurement
on a blood sample from a
healthy donor.
- Contact Radiometer
service representative.
293
Warning:
HbF
detected and
compensat-
ed for
Detected FHbF is
greater than 20 % of
ctHb. Oxi parameters
are automatically
corrected.
No action required.
For information only.
326
Empty Gas
1 cylinder
Gas 1 cylinder is
detected as empty.
327
Empty Gas
2 cylinder
Gas 2 cylinder is
detected as empty.
- Enter Replacement mode
via Replacement Status.
- Scan the barcode of a new
Gas 1 or 2 cylinder.
- Replace Gas 1 cylinder.
(See chapter 7:
Replacements for detailed
information.)
Removal condition: Gas
pressure in Gas 1 or 2
cylinder.
Continued on next page
11-14