ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
328 No leading
air segment
in inlet's
liquid
sensor
within time
frame
Air segment in front of
sample is too small or
missing during sample
aspiration, possibly due
to obstructions in inlet
or problems with liquid
sensors. Measurement
is aborted.
- Check inlet for
obstructions and clean as
required.
- Perform liquid sensor
adjustment.
- Perform the Fluid
Transport Troubleshooting
procedure.
329 QC
expiration
date
exceeded
The quality control
measurement was
performed on an
expired control
solution.
Discontinue the use of the
lot and set up a valid lot for
the control solution.
331 No sample
detected
during
sample
aspiration
No sample detected in
the pH/BG module.
Measurement is
aborted.
- Ensure that adequate
sample volume is used.
- Check the sample for
clots.
- Perform the Fluid
Transport Troubleshooting
procedure.
332,
333
pH/BG
module not
filled
Sample did not fill the
pH/BG module
properly during
aspiration.
Measurement is
aborted.
- Ensure that adequate
sample volume is used.
- Check the sample for
clots.
- Perform the Fluid
Transport Troubleshooting
procedure.
339-340.
342-343,
345-346
Pump
calibration
error
The pump calibration
failed for one or more
pumps. The previous
pump calibration is
used. A successful
pump calibration
should be attempted as
soon as possible.
- Check rinse solution for
sufficient volume. Replace
as necessary.
- Repeat pump calibration.
- Perform the Pump
Troubleshooting procedure.
Removal condition:
Successful pump
calibration.
Continued on next page