ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
481-482 Conditioning
error
Indicates a fluid
transport problem of
pH/BG module during
Flush.
484 Today is last
day in
current
statistical
month –
remember to
print QC
statistics
After the current day,
quality control
statistics obtained over
the month will be
deleted and new
statistics started.
Print the QC statistics if a
copy is required.
487 A new
statistical
month has
begun –
remember to
export WDC
data
A new statistical
month has begun.
Make a WDC report disk.
Removal condition: A
WDC report disk has been
made.
493 Warning:
Bilirubin
detected and
compensat-
ed for
Detected bilirubin
concentration is
greater than
50 umol/L. The
corresponding plasma
bilirubin concentration
can be calculated as
follows: ctBil(blood) =
(1-Hct) x
ctBil(plasma).
No action required.
494 Bilirubin too
high
Detected bilirubin
concentration,
ctBil(blood), is greater
than 2000umol/L. The
corresponding plasma
bilirubin concentration
can be calculated as
follows: ctBil(blood) =
(1-Hct) x
ctBil(plasma).
No action required.
Continued on next page