11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
499 Sample too
early in Oxi
liquid
sensor.
The Oxi module
detected liquid
prematurely during
sample aspiration. The
activity is aborted.
- Check the sample for
clots.
- Repeat measurement.
- Replace inlet gasket.
- Perform the Fluid
transport troubleshooting
procedure.
500 Rinse
detected by
pH/BG
module
when not
expected
The pH/BG module
detected liquid
unexpectedly.
Perform the Fluid transport
troubleshooting procedure.
501 Rinse not
aspirated
into the
pH/BG
module
within time
limit
The pH/BG module
was not filled properly
with Rinse Solution.
- Check the Rinse Solution
for sufficient volume and
proper mounting. Replace
as necessary.
- Perform the Fluid
transport troubleshooting
procedure.
503 Unable to
fill inlet
with
continuous
rinse
segment
during
Pump
Calibration
Three attempts to fill
the inlet probe with a
continuous rinse
segment have failed;
the Pump Calibration
program aborts.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform the Pump
troubleshooting procedure.
504 No sample
in El/Met
lower liquid
sensor
within time
limit
The sample did not fill
El/Met module
properly. The activity is
aborted.
- Check the sample for
clots.
- Ensure that sample
volume matches selected
sample mode.
- Perform the Fluid
transport troubleshooting
procedure.
Continued on next page
11-22