ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
505 Met II upper
liquid
sensor did
not detect
Cleaning
Solution
The Met II module was
not filled properly
during Cleaning
program. The activity
is aborted.
506 El/Met
upper liquid
sensor did
not detect
Cleaning
Solution
The El/Met module
was not filled properly
during Cleaning
program. The activity
is aborted.
507 pH/BG
upper liquid
sensor did
not detect
Cleaning
solution
The pH/BG module
was not filled properly
during Cleaning
program. Cleaning is
aborted.
- Check the Cleaning
solution for sufficient
volume. Replace as
required.
- Perform Fluid transport
troubleshooting procedure.
Removal condition.
Successful Cleaning.
508 Rinse error Rinse Solution was not
detected during Rinse
program.
Measurements and
calibrations cannot be
performed until the
condition is removed.
- Check the rinse solution
for sufficient volume.
Replace as required.
- If the message "Inlet flow
impeded" (765) has been
reported, replace or clean
inlet probe.
- Perform Fluid transport
troubleshooting procedure.
Removal condition:
Successful Rinse.
509 pH/BG
upper liquid
sensor did
not receive
Rinse from
above
The pH/BG module
was not filled properly
during Rinse program;
automatic Refill is
attempted.
510 El/Met
upper liquid
sensor did
not receive
Rinse from
above
The El/Met module
was not filled properly
during Rinse program;
automatic Refill is
attempted.
- Check the Rinse Solution
for sufficient volume.
Replace as required.
- Perform Fluid transport
troubleshooting procedure.
Removal condition:
Successful Rinse.
Continued on next page
11-23