11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
512 Temperatu-
re error
The temperature was
outside the required
range during
measurement or
calibration.
All results are marked
with "?".
Ensure ambient
temperature is between 15
and 32 °C.
- If the analyzer has
recently performed a Cold
start, wait for temperature
error to disappear.
- Shield analyzer from
direct sunlight or heat
sources.
- Contact Radiometer
service representative.
513 No gas Measurement or
calibration was
performed with Gas
error present.
Activity is aborted.
- Check preasure in gas
cylinders. Replace as
required.
- Verify that cylinders are
properly installed.
- Enter the Replacement
screen (via Replacement
Status) and re-scan the
barcode labels of the gas
cylinders.
521 Inhomoge-
neous
sample
Air bubbles were
detected in the sample.
Results may have "?".
- Ensure sampler and
sample volume match the
selected sample mode.
- Repeat measurement.
- Check pump tubes and
inlet for leaks. Replace
tubing and inlet gasket as
required.
529, 531-
532, 534-
537
Liquid
sensor
calibration
error
One or more of the
liquid sensors failed to
calibrate.
- Repeat liquid sensor
calibration.
- Perform Fluid transport
troubleshooting procedure.
- Contact Radiometer
Service Representative.
Continued on next page
11-24