ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
538 pH/BG
pump error
The pH/BG module
pump failed to
calibrate. Previous
calibration is used until
a successful pump
calibration is
performed.
- Perform Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
539 El/Met
pump error
The El/Met module
pump failed to calibra-
te. Previous calibration
is used until a success-
ful pump calibration is
performed.
- Perform Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
540 Met II pump
error
The Met II module
pump failed to
calibrate. Previous
calibration is used until
a successful pump
calibration is
performed.
- Perform Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
541 Solution
pump error
The solution pump
failed to calibrate.
Previous calibration is
used until a successful
pump calibration is
performed.
- Perform Pump
Troubleshooting procedure.
- Contact Radiometer
service representative.
564 No
Cleaning
Solution
detected by
Oxi liquid
sensor
during
initializing.
No Cleaning Solution
was detected by Oxi
module during Startup.
- Check the Cleaning
solution for sufficient
volume. Replace as
necessary.
- Perform Fluid transport
troubleshooting procedure.
567 No Cal 2
Solution
detected by
Oxi liquid
sensor
during
initializing
No Cal 2 Solution was
detected by Oxi module
during Startup.
- Check the Cal 2 solution
for sufficient volume.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
Continued on next page
11-25