11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
569 No Cal 1
Solution
detected by
Oxi liquid
sensor
during
initializing
No Cal 1 Solution was
detected by the Oxi
module during Startup.
Check the Cal 1 solution
for sufficient volume.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
571 No Rinse
Solution
detected by
OXI liquid
sensor
during
initializing
No Rinse Solution was
detected by the OXI
module during Startup.
- Check the Rinse solution
for sufficient volume.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
573 Oximetry
hardware
error
The OXI module has a
hardware error.
- Perform a 2-point
calibration.
- Contact Radiometer
Service Representative if
the problem persists.
574 Oximetry
calibration
error
The measuring system
has no calibration data.
It is not possible to
make a measurement.
- Perform a 2-point
calibration.
- Contact Radiometer
Service Representative.
Removal condition:
Successful 2-point
calibration.
577-578 Option key
error
An error occurred in
the analyzer protection
system. The analyzer
cannot be used.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart it.
- Contact Radiometer
Service Representative.
Removal condition:
Key installed correctly and
accepted by analyzer.
579 Oximetry
hardware
problem.
Not possible
to measure.
Parameter cannot be
realiably measured
because of a hardware
problem.
Perform a 2-point
calibration.
- Contact Radiometer
service representative.
Removal condition: No
errors detected by oximetry
hardware.
Continued on next page
11-26