ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
580 Oximetry
hardware
problem.
Possible to
measure.
A hardware problem
exists in the OXI
module. Parameter can
be measured.
- Perform a 2-point
calibration.
- Contact Radiometer
service representative.
Removal condition: No
errors detected by oximetry
hardware.
581 Oximetry
measuring
error
Spectrum deviates from
expected blood or QC
spectrum.
Measurement may be
unrealiable.
- Check patient record for
medication containing
possible interfering
substances.
- Call a 2-point calibration.
- Contact Radiometer
Service Representative.
582 tHb
calibration
outside
limits
tHb calibration failed. Repeat tHb calibration.
Removal condition:
Successful tHb calibration.
583 Measured
value
outside
reportable
range for
the
parameter
The parameter value is
outside the defined
reportable range or the
analyzer measurement
range.
- Check for and remedy
other errors in the result,
and any errors in System
Messages or Calibration
Status.
- Perform QC. If the QC
result value lies within the
measuring range, the blood
sample may be suspected.
- Perform the Electrode
Troubleshooting procedure.
588 Measured
QC value
lower than
statistical
range
The parameter value is
below the lower limit
of the user-defined
statistical range.
Measurement is not
included in statistics.
- Verify procedure and
repeat the measurement.
- Refer to the Quality
Control Systems Reference
Manual for details on the
evaluation of the results.