ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
605 Cal expired
606 Cal expired
(pH)
608 Cal expired
(pCO
2
)
609 Cal expired
(pO
2
)
610 Cal expired
(K)
611 Cal expired
(Na)
612 Cal expired
(Ca)
613 Cal expired
(Cl)
614 Cal expired
(Glu)
615
Cal expired
(Lac)
616 Cal expired
(OXI)
Too long time passed
since the last successful
calibration of the
parameter. Calibrations
of the parameter are
considered invalid and
values reported as
".....".
Perform 1- or 2-point
calibration.
Removal condition:
Successful 1- or 2-point
calibration.
618 Waste
detector
error
The waste detector may
be faulty or damaged.
Contact Radiometer service
representative.
Removal condition:
Functioning waste detector
installed.
619 Waste
container
missing
The waste container is
missing or improperly
mounted.
Mount the waste container
properly.
Removal condition: Waste
container properly
mounted.
Continued on next page
11-29