11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
620 Waste
container
full
Waste container is
filled to near-maximum
capacity. Unless
replaced soon, the
analyzer will enter
Standby until the
container is replaced.
Replace waste container.
Removal condition: Empty
Waste container installed.
621 Waste
container
critically
full
The waste container is
filled to maximum
capacity. The analyzer
enters Standby mode;
no measurements or
calibrations can be
performed.
Replace waste container.
Removal condition: Empty
Waste container installed.
622
Expected
liquid level
below 0
One or more of the
solution barcodes was
entered incorrectly
during replacement.
The analyzer cannot
monitor solution
consumption.
Check the solution
containers and replace as
necessary. Remember to
scan barcodes.
Removal condition:
Solution containers
correctly installed and
solution barcodes correctly
recorded by the analyzer.
623 Solution
empty
Wet section activities
are suspended due to
one or more empty
solution containers.
Check the solution levels
and replace containers as
necessary.
Removal condition:
Sufficient level of all
solutions.
624 Solution
missing
One or more solution
containers is missing or
improperly mounted.
Check that all solution
containers are mounted
properly.
Removal condition: All
solution containers are
mounted correctly.
Continued on next page
11-30