ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
628 No
Cleaning
Solution
detected by
Oxi liquid
sensor
during
Refill
OXI module was not
filled correctly with
Cleaning Solution.
- Check Cleaning solution
level in the container.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
630 No Cal 2
Solution
detected by
Oxi liquid
sensor
during
Refill
OXI module was not
filled correctly with Cal
2 Solution.
- Check Cal 2 Solution
level in the container.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
632 No Cal 1
Solution
detected by
Oxi liquid
sensor
during
Refill
OXI module was not
filled with Cal 1
Solution.
- Check Cal 1 Solution
level in the container.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
634 No Rinse
Solution
detected by
Oxi liquid
sensor
during
Refill
OXI module was not
filled with Rinse
Solution.
- Check Rinse Solution
level in the container.
Replace as necessary.
- Perform Fluid transport
troubleshooting procedure.
636
Gas 1
cylinder low
pressure
Low pressure level is
detected in Gas 1
cylinder.
Replace the Gas 1 cylinder
within a few days.
Removal condition:
Sufficient pressure in Gas 1
cylinder.
637
Gas 2
cylinder low
pressure
Low pressure level is
detected in Gas 2
cylinder.
Replace the Gas 2 cylinder
within a few days.
Removal condition:
Sufficient pressure in Gas 2
cylinder.