ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
654 Sample in
El/Met
upper liquid
sensor
detected
prematurely
during
aspiration
El/Met module was not
filled correctly during
sample aspiration.
- Ensure sample is free of
clots.
- Repeat measurement.
- Perform Fluid transport
troubleshooting procedure.
655 Sample in
pH/BG
upper liquid
sensor
detected
prematurely
during
aspiration
pH/BG module was not
filled correctly during
sample aspiration.
-Ensure sample is free of
clots.
- Repeat measurement.
- Perform Fluid transport
troubleshooting procedure.
656 Sample in
OXI liquid
sensor
detected
prematurely
during
aspiration
OXI module was not
filled correctly during
sample aspriration.
- Ensure sample is free of
clots.
- Repeat measurement.
- Perform the Fluid
Transport Troubleshooting
procedure.
658 Inlet flaps
removed
The analyzer entered
HOLD because both
inlet flaps were
removed.
Ensure the inlet flaps are
properly mounted prior to
restart.
Removal condition: Inlet
flaps properly mounted.
659 Cover
removed or
window
opened
The analyzer entered
HOLD because the
cover was removed or
the window to the
electrode modules
opened.
Position the cover and
close the window prior to
restart.
Removal condition: Cover
in position and window
closed.
660 Inlet
calibration
error
An error occurred
during Inlet calibration.
Analyzer entered
HOLD mode.
- Remove the inlet gasket. -
Inspect the gasket and inlet
probe. Replace as
necessary.
- Press Restart.
Removal condition:
Successful Inlet calibration
Continued on next page
11-33