11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
661 Leak
detected
A leak was detected.
The analyzer entered
HOLD mode.
- Remove the analyzer
covers and inspect the fluid
transport system for leaks.
- Check the inlet gasket,
pump tubes and ensure
electrodes are mounted
correctly. Repair as
necessary
- Restart the analyzer.
During restart the analyzer
performs a leak detection.
Removal condition: Leak
test completed without any
leaks detected.
662 Barometer
out of range
Measured barometer
value is outside the
measuring range: 60-
106.7 kPa.
Contact Radiometer service
representative.
663 Leakage A leak was detected in
the fluid transport
system.
Remove the analyzer
covers and inspect the fluid
transport system for leaks.
- Check the inlet gasket,
pump tubes and ensure
electrodes are mounted
correctly. Repair as
necessary.
- Restart the analyzer.
During restart the analyzer
performs a leak detection.
Removal condition: Leak
test completed without any
leaks detected.
664 Sample
problem
A problem was
detected during sample
transport through the
analyzer.
- Check the sample for
clots.
- Ensure sample volume
matches selected sample
mode.
- Perform the Fluid
Transport Troubleshooting
procedure.
Continued on next page
11-34