11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
680 pH/BG
module not
active
The pH/BG module is
not responding due to
internal communication
problem.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart it.
- Contact Radiometer
Service Representative.
Removal condition:
pH/BG module ready.
681
El/Met
module not
active
The El/Met module is
not responding due to
internal communication
problem, or software
configuration does not
match the analyzer
type.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart it.
- Contact Radiometer
Service Representative.
Removal condition: El/Met
module ready, or software
configured without El/Met
module support.
682 Oxi module
not active
The Oxi module is not
responding due to
internal communication
problem, or software
configuration does not
match the analyzer
type.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart it.
- Contact Radiometer
Service Representative.
Removal condition: OXI
module ready, or software
configured without OXI
module support.
683 Inlet
module not
active
The inlet module is not
responding due to
internal communication
problem.
- Shut down power using
the Temporary Shutdown
function, then restart the
analyzer.
- Contact Radiometer
Service Representative.
Removal condition: Inlet
module ready.
Continued on next page
11-36