ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
688 ctHb/ceHb
too low for
OXI
calculation
ctHb < 1 mmol/L or
ceHb < 0.75 mmol/L. If
ctHb is too low FHHb,
FO
2
Hb, FCOHb and
FmetHb are not
calculated. If ceHb =
cHHb + cO
2
Hb is too
low sO
2
is not
calculated.
Informative message. No
analyzer error detected.
689
Gas
pressure
offset
voltage out
of range.
The gas pressure offset
voltage measured by
the transducer is out of
range: 0.07-0.35 V.
- Restart analyzer.
- Contact Radiometer
service representative.
Removal condition: Gas
pressure transducer offset
within the required range.
692 ABL not
connected to
RADIANCE
The analyzer is not
connected to
RADIANCE.
Contact your RADIANCE
/IT engineer.
- Check RADIANCE
communication setup
including TCP/IP address,
port No. and password.
- Check that RADIANCE
is responding.
- Check network
connections.
Removal condition:
RADIANCE connection
established or disabled.
693 ABL not
connected to
RADIANCE
- incorrect
password.
The analyzer was
refused connection to
RADIANCE due to
incorrect password.
Enter the correct password
in the analyzer's
RADIANCE
Communication setup.
Removal condition:
RADIANCE connection
established or disabled.
694 ABL
connected to
RADIANCE
The analyzer is
connected to
RADIANCE.
No action required.