ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
705 AutoCheck
QC
measure-
ment is
repeated
twice
The scheduled QC
measurement was not
accepted; the
measurement was
repeated twice as
requested in Setup
program: Corrective
Actions.
No action required.
707 Replace-
ment
overdue by
10 %.
Analyzer
locked
Replacement is
overdue by 10%
(corrective action
"Lock analyzer" was
selected in Setup
program: Corrective
Actions). When the
analyzer is locked,
scheduled calibrations
are performed, but no
patient samples or QC
measurements are
allowed.
- Check Replacement
Status and replace as
required.
- Unlock analyzer in
Miscellaneous Setup
program.
Removal condition: No
replacement pending.
708 Corrective
action not
possible due
to empty
AutoCheck
carousel
Scheduled AutoCheck
measurement was
requested, but the
carousel was empty.
Refill AutoCheck carousel.
709
El/Met
module not
expected
Software configuration
does not match
analyzer type.
Contact Radiometer service
representative.
Removal condition:
Matching software and
hardware configuration.
710
OXI module
not
expected.
Software configuration
does not match
analyzer type.
Contact Radiometer service
representative.
Removal condition:
Matching software and
hardware configuration.
711
Warning:
DysHb high
Blood sample has a
high level of DysHb.
FCOHb + FMetHb >
5%.
Informative message. No
analyzer error was
detected.
Continued on next page
11-39