11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
712 FHbF
measure-
ment not
possible
Composition of blood
sample makes FHbF
measurement too
inaccurate, but OXI
parameters are
corrected for HbF. See
explanation in the
Reference Manual.
Informative message. No
analyzer error was
detected.
713 ctBil
measure-
ment not
possible
Blood sample ctHb is
so high that nearly no
plasma is left to
measure plasma
bilirubin on.
ctHb > 15.5mmol/L.
Informative message. No
analyzer error was
detected.
714,
740-743
AutoCheck
mechanical
error
A mechanical error
occurred in one of the
AutoCheck sub-
modules.
- Check the AutoCheck
module cover. Close if
opened.
- If error persists, contact
Radiometer service
representative.
715 AutoCheck
communica-
tion error
Communication error
between AutoCheck
submodules.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart the analyzer.
- Contact Radiometer
Service Representative.
716 AutoCheck
cover
missing
The cover was removed
from the AutoCheck
module.
Mount the AutoCheck
cover.
Removal condition:
AutoCheck cover is
mounted correctly.
717 Carousel
packing not
optimal
The AutoCheck
carousel was not
packed according to the
schedule.
Repack AutoCheck
carousel according to the
packing list.
Removal condition:
AutoCheck carousel
packed according to
Packing list.
Continued on next page