11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
725 No Cal 1
Solution
detected by
inlet during
Startup
pH/BG module was not
filled correctly with Cal
1 Solution during
Startup.
- Check Cal 1 solution
level. Replace container as
necessary.
- Perform Fluid transport
troubleshooting procedure.
726 No Rinse
Solution
detected by
inlet during
Startup
pH/BG module was not
filled correctly with
Rinse Solution during
Startup.
- Check Rinse solution
level. Replace container as
necessary.
- Perform Fluid transport
troubleshooting procedure.
727 No Cleaning
Solution
detected by
inlet during
Refill
pH/BG module was not
filled correctly with
Cleaning Solution
during Refill.
- Check Cleaning solution
level. Replace container as
necessary.
- Perform Fluid transport
troubleshooting procedure.
728 No Cal 2
Solution
detected by
inlet during
Refill
pH/BG module was not
filled correctly with Cal
2 Solution during
Refill.
- Check Cal 2 solution
level. Replace container as
necessary.
- Perform Fluid transport
troubleshooting procedure.
729 No Cal 1
Solution
detected by
inlet during
Refill
pH/BG module was not
filled correctly with Cal
1 Solution during
Refill.
- Check Cal 1 solution
level. Replace container as
necessary.
- Perform Fluid transport
troubleshooting procedure.
730 No Rinse
Solution
detected by
inlet during
Refill
pH/BG module was not
filled correctly with
Rinse Solution during
Refill.
- Check Rinse solution
level. Replace container as
necessary.
- Perform Fluid transport
troubleshooting procedure.
731 Met II upper
liquid
sensor did
not receive
Rinse from
above
Met II module was not
filled correctly during
Rinse program;
automatic Refill is
attempted.
- Check the Rinse Solution
for sufficient volume.
Replace as required.
- Perform Fluid transport
troubleshooting procedure.