ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
734
DMS not
connected to
Analyzing
Unit
Data management
system establishes
connection to the
Analyzing Unit, or the
connection is lost.
- Wait a few minutes for
the connection to establish.
- Restart the analyzer.
- If the error persists,
contact Radiometer service
representative.
738 Ampoule(s)
in
AutoCheck
carousel not
installed
AutoCheck carousel
contains an ampoule
which was not selected
in Control Solutions
Setup program.
Read in the barcodes for all
ampoules present in
AutoCheck carousel – use
inserts.
Removal condition:
Barcodes read in for all
ampoules present in the
AutoCheck carousel.
739 Solution's
composition
not
identifiable
AutoCheck ampoule
solutions do not match
labeling.
- Repeat measurement.
- Contact Radiometer
service representative if the
problem persists.
744 AutoCheck
Laser Setup
Error
AutoCheck barcode
reader has failed.
- Restart analyzer.
- If the error persists,
contact Radiometer service
representative.
745 Low hard
disk space
Free hard disk space is
below 100 Mb.
Move archive files to
another storage device.
Removal condition: More
than 100 Mb of free hard
disk space.
746 Rinse
Solution
low warning
Rinse Solution level is
below the defined
warning level.
Replace container as
required.
Removal condition:
Solution level above
warning level.
Continued on next page
11-43