11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
747
Cal 1
solution low
warning
Cal 1 Solution level is
below the defined
warning level.
Replace container as
required.
Removal condition:
Solution level above
warning level.
748
Cal 2
Solution
low warning
Cal 2 Solution level is
below the defined
warning level.
Replace container as
required.
Removal condition:
Solution level above
warning level.
749
Cleaning
Solution
low warning
Cleaning Solution level
is below the defined
warning level.
Replace container as
required.
Removal condition:
Solution level above
warning level.
750 Solution
projection
adjusted
The estimated solution
level was adjusted by
the operator.
No action required.
751
Message 751 is only found in the Activity Log to inform the user
about activities that have taken place.
The message is blank (empty) in the database, and when an activity
occurs the actual status information is appended to it resulting in
the logged 751-message.
If the setting "Log All Measuring Activities" is enabled in
Miscellaneous Setup, all wet section activities will be logged in the
Activity Log as 751-messages.
753 Service
Electrode
Updatings
enabled
Logging of sensor
signals was enabled.
This may be due to
ongoing technical
service activities on the
analyzer. The condition
does not affect analyzer
functions.
Contact Radiometer service
representative if logging
was not enabled on
purpose.
Removal condition:
Electrode signal logging
disabled by Radiometer
service representative.