ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
763
AutoCheck
temperature
out of
required
range
The QC temperature
was outside the range
valid for the
temperature correction
algorithms.
- Ensure the ambient
temperature is between 15
and 32 degrees Celsius.
- If the system has just
performed a cold start, wait
for error to disappear.
- Replace fan filter if dirty.
- Shield analyzer from
direct sunlight and other
heat sources.
- Contact Radiometer
Service Representative.
Removal condition:
AutoCheck temperature
within the required range.
764 Liquid
detected in
pH/BG
module
during
Expired Air
sample
Liquid sample detected
during aspiration in the
Expired Air mode.
Do not use the Expired Air
mode for liquid samples.
765
Inlet flow
impeded
Inlet flow is impeded
by some foreign matter.
- Perform Inlet
troubleshooting procedure.
- Perform the Protein
Removal program.
Removal condition: Inlet
has no obstructions.
Continued on next page