11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
766 ABL not
connected to
RADIANCE
– no
RADIANCE
connection
license
The analyzer has been
refused connection to
RADIANCE because
there is no connection
licence available on
RADIANCE.
Contact RADIANCE /IT
engineer or Radiometer
service representative.
Removal condition:
Connection to RADIANCE
established.
767 ABL not
connected to
RADIANCE
– ABL
StatLink
version too
high
The analyzer has been
refused connection to
RADIANCE because
the ABL StatLink
version is higher than
the RADIANCE
StatLink version.
Contact RADIANCE /IT
engineer or Radiometer
service representative.
Removal condition:
RADIANCE connection
established.
768 ABL not
connected to
RADIANCE
– ABL
StatLink
version too
low
The analyzer has been
refused connection to
RADIANCE because
the ABL StatLink
version is lower than
the RADIANCE
StatLink version.
Contact RADIANCE /IT
engineer or Radiometer
service representative.
Removal condition:
RADIANCE connection
established.
770 Failed to
restore
Custom
Setup
The setup could not be
restored.
- Download the setup data
from another floppy disk,
hard disk or network.
- Contact Radiometer
service representative if the
error persists.
771 Succeeded to
restore
Custom
Setup
Restoring of setup is
completed.
No action required.
772 User
Activity:
User activity logged by
operator.
No action required.
773 Remote
operator
logged on
with user:
A remote operator has
logged on the analyzer
via NetOp
No action required.
Continued on next page
11-46