ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
790 Adaptive
measuring
mode
applied
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
sample handling in
order to report as many
of the desired
parameters as possible.
If missing parameters are
needed, repeat the
measurement with
sufficient sample free of
clots and air bubbles.
791 Parameter
could not be
measured
Due to insufficient or
inhomogenous sample,
the parameter in
question could not be
measured.
If missing parameters are
needed, repeat the
measurement with
sufficient sample free of
clots and air bubbles.
792 Liquid
sensor
calibration
error
Inlet lower liquid
sensor failed to
calibrate.
- Repeat liquid sensor
calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
- Contact Radiometer
service representative.
793 pH/BG
pump
asymmetry
error
The pH/BG module
pump failed to calibrate
pump asymmetry. The
same calibration values
will be used for both
directions until a
successful calibration
of pump asymmetry is
performed.
- Perform the Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
794 El/Met
pump
asymmetry
error
The El/Met module
pump failed to calibrate
pump asymmetry. The
same calibration values
will be used for both
directions until a
successful calibration
of pump asymmetry is
performed.
- Perform the Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
Continued on next page
11-49