11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
795 Met II pump
asymmetry
error
The Met II module
pump failed to calibrate
pump asymmetry. The
same calibration values
will be used for both
directions until a
successful calibration
of pump asymmetry
has been performed.
- Perform the Pump
troubleshooting procedure.
- Contact Radiometer
service representative.
796 Solution
pump
asymmetry
error
The Solution module
pump failed to calibrate
pump asymmetry. The
same calibration values
will be used for both
directions until a
successful calibration
of pump asymmetry is
performed.
- Perform the Pump
Troubleshooting procedure.
- Contact Radiometer
service representative.
797 Waste pump
asymmetry
error
The Waste pump in the
Solution module failed
to calibrate pump
asymmetry. The same
calibration values will
be used for both
directions until a
successful calibration
of pump asymmetry is
performed.
- Perform the Pump
Troubleshooting procedure.
- Contact Radiometer
service representative.
798 User logged
on
User logged on
successfully.
No action required.
Informative message.
799 User logged
off
User logged off. Informative message. No
action required.
800 Logon
attempt
failed
User tried to log on but
did not provide a valid
password.
Provide a valid password to
log on.
801 Hard disk
S.M.A.R.T.
threshold(s)
exceeded
The analyzer's hard
disk drive has given a
warning. This may
mean imminent hard
disk drive failure.
Contact Radiometer service
representative.
Continued on next page
11-50