ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
805 Adaptive
measuring
mode appli-
ed (decay to
pH/BG/Met/
Oxi)
806 Adaptive
measuring
mode appli-
ed (decay to
all micro
mode)
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
sample handling in
order to report as many
of the desired
parameters as possible.
- If missing parameters are
needed, repeat the
measurement with
sufficient sample free from
clots and air bubbles.
807 Unautho-
rized
software
Software option key is
not installed. Analyzer
cannot be used.
Contact Radiometer service
representative.
808 Calibration
drift not
available
A calibration drift
value could not be
calculated, since this is
the first calibration
result after calibration
drift monitoring was
started.
This message may
appear on the first
calibration result after
the analyzer has been
turned on or after the
parameter in question
has been enabled.
A drift value will be
calculated for the next
calibration result.
No action required. The
message is informative.
809 Service
Electrode
Updatings
disabled
The message is used in
the Activity Log to
indicate that Service
Electrode Updatings
have been disabled.
No action required. The
message is for information
only.