11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
810-827
829, 830
Parameter
locked (pH,
pCO
2
, pO
2
,
cK
+
, etc.)
Parameter has been
locked by a
RADIANCE operator,
as reflected in the
Activity Log.
When a parameter is
locked, presumably due
to problems with QC,
the parameter is repres-
sed in patient results.
Await corrective actions
initiated by the
RADIANCE operator.
Removal condition:
Determined by the
RADIANCE operator.
831-848,
850, 851
Parameter
unlocked
(pH, pCO
2
,
pO
2
, cK
+
,
etc.)
The message is used in
the Activity Log to
indicate that a
previously locked
parameter has been
unlocked.
Informative message. No
action required.
852 Radiance: Message from
RADIANCE
Informative message. No
action required.
853 LIS/HIS:
Missing
Communi-
cation
Option key
No valid option key for
LIS/HIS
communication is
installed on the
analyzer.
Contact Radiometer service
representative.
854 No leading
air segment
in inlet's
lower liquid
sensor
within time
frame
Air segment in front of
sample is too small or
missing during sample
aspiration, possibly due
to obstructions in inlet
or problems with liquid
sensor. Measurement is
aborted.
- Check inlet for
obstructions and clean as
required.
- Perform liquid sensor
adjustment.
- Perform the Fluid
Transport Troubleshooting
procedure.
855 Base Excess
out of range
Base Excess exceeds
+/- 30 mmol/L range
Informative message. No
analyzer error was
detected.
Continued on next page
11-52