ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
862 pH-only
sample
handling
adapted to
35uL range
864 pH-only
sample
handling
adapted to
55uL range
To counteract
insufficient or
inhomogeneous
sample, the analyzer
has automatically
adjusted the sample
handling to the sample
volume.
None.
865 pH/BG
sample
handling
adapted to
35uL range
(decay to
pH only)
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
measuring mode in
order to report as many
of the desired
parameters as possible.
If missing parameters are
needed, repeat the
measurement with
sufficient sample free from
clots and air bubbles.
867 pH/BG
sample
handling
adapted to
55uL range
868 pH/BG/Oxi
sample
handling
adapted to
55uL range
To counteract
insufficient or
inhomogeneous
sample, the analyzer
has automatically
adjusted the sample
handling to the sample
volume.
None.
Continued on next page