11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
869 Adaptive
sample
handling
applied
(decay to
pH/Oxi
without Bil)
870 Adaptive
sample
handling
applied
(decay to
pH/Oxi)
871 Adaptive
sample
handling
applied
(decay to
pH/BG/Oxi)
872 Adaptive
measuring
mode
applied
(pH/BG/Oxi
[2])
873 Adaptive
measuring
mode
applied
(pH/BG/
Met/Oxi)
874 Adaptive
measuring
mode
applied
(decay to all
micro
mode)
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
measuring mode in
order to report as many
of the desired
parameters as possible.
If missing parameters are
needed, repeat the
measurement with
sufficient sample free from
clots and air bubbles.
875 Sample
aged
The specified limit for
sample age has been
exceeded.
Draw and analyze new
sample.
Continued on next page