ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
876 FLEXQ
positioning
warning
A problem was
detected during
positioning of the
sampler tray
The message is informative
and intended for service
use only
Removal condition:
No FLEXQ positioning
problems observed for 24
hours.
877 FLEXQ
positioning
error
A sampler in the
sampler tray could not
be aligned with the
syringe inlet; aspiration
of the sample was not
possible.
- Remove all the samplers
from the sampler tray (the
sampler registration is
lost).
- If the sampler tray is
blocking the syringe inlet
or AutoCheck module, try
to push it gently to its
home position.
- If the syringe inlet is not
blocked, scan the sampler
barcode, using the
analyzer’s barcode reader
and perform a manual
measurement.
- To get the FLEXQ
operational again, restart
the analyzer, using
Temporary Shutdown and
power off/on.
- If the error persists,
contact Radiometer service
representative.
- Removal condition:
Successful FLEXQ
sampler tray calibration.
Continued on next page
11-55