11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
878 No Rinse
solution
during
calibration
The analyzer was not
able to retrieve Rinse
solution during
calibration.
Check level in Rinse
Solution bottle. Replace
bottle as necessary.
879 No Cal1
solution
during
calibration
The analyzer was
unable to retrieve Cal1
solution during
calibration.
Check level in Cal1
solution bottle. Replace
bottle as necessary.
880 No Cal2
solution
during
calibration
The analyzer was
unable to retrieve Cal2
solution during
calibration.
Check level in Cal2
solution bottle. Replace
bottle as necessary.
881 LIS/HIS:
Missing
CIC Com-
munication
Option Key
No valid Option Key
for CIC communication
is installed on the
analyzer.
Contact Radiometer service
representative.
882 Mixer tray
slip detected
A small positioning
problem of the mixer
tray has been detected
and corrected.
- Lift the sampler tray
cover and check for
obstructions in the mixer
tray area.
- Remove any foreign
objects.
- Remount the sampler tray
cover.
Removal condition:
Alignment operation
performed without slip.
883 Mixer error The sample may not
have been adequately
mixed.
- Lift the sampler tray
cover and check for
obstructions in the mixer
tray area.
- Remove any foreign
objects.
- Remount the sampler tray
cover.
Continued on next page
11-56