11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
890 Error
measuring
Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
892 Inlet not
filled
properly
during
pump
calibration
using
reagent as
backup
During pump
calibration
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
893 Sample in
Met II's
upper liquid
sensor
detected
prematurely
during
aspiration
Met II module was not
filled correctly during
sample aspiration.
- Ensure sample is free of
clots.
- Repeat measurement.
- Perform the Fluid
Transport Troubleshooting
procedure.
Continued on next page
11-58