ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
894 Adaptive
measuring
mode
applied
(pH/BG/
Met/Oxi
[2])
895 Adaptive
measuring
mode
applied
(pH/BG/
Met/El/Oxi)
896 Adaptive
measuring
mode
applied
(pH/BG/
Met/El/Oxi
[2])
897 Adaptive
measuring
mode
applied
(pH/BG/
Met/El/Oxi
[3])
898 Adaptive
measuring
mode
applied
(pH/BG/
Met/El/Crea
/Oxi)
899 Adaptive
measuring
mode
applied
(pH/BG/
Met/El/
Crea/Oxi
[2])
Due to insufficient or
inhomogeneous
sample, the analyzer
has automatically
chosen an adaptive
measuring mode in
order to report as many
of the desired
parameters as possible.
If missing parameters are
needed, repeat the
measurement with
sufficient sample free from
clots and air bubbles.
Continued on next page
11-59

Table of Contents