ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
920-922 Module not
filled within
time limit in
wash with
cleaning
Module did not fill
properly with cleaning
solution.
- Check the solution
containers for adequate
volume. Replace as
required.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful Rinse
923 Inlet not
refilled
properly
during bi-
directional
pump
calibration
The inlet module was
not filled properly
during bi-directional
pump calibration.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
924 Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 1
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
925 Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 2
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure
Removal
condition:Successful pump
calibration.
Continued on next page
11-61