EasyManua.ls Logo

RADIOMETER ABL800 FLEX - Page 343

RADIOMETER ABL800 FLEX
438 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
920-922 Module not
filled within
time limit in
wash with
cleaning
Module did not fill
properly with cleaning
solution.
- Check the solution
containers for adequate
volume. Replace as
required.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful Rinse
923 Inlet not
refilled
properly
during bi-
directional
pump
calibration
The inlet module was
not filled properly
during bi-directional
pump calibration.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
924 Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 1
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
925 Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 2
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure
Removal
condition:Successful pump
calibration.
Continued on next page
11-61

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Related product manuals