11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
926 Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 3
927 Inlet's liquid
sensor to
Oxi's liquid
sensor
volume
could not be
confirmed 4
928 Unable to
fill Inlet
with
continuous
rinse during
pump
calibration
During pump
calibration,
measurement of
internal volume part
failed.
Causes the Fluid
Transport Control
program to abort.
- Check Rinse Solution for
sufficient volume. Replace
as necessary.
- Perform Pump
Calibration.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Successful pump
calibration.
929 TIPCAP
conflict
detected 1
930 TIPCAP
conflict
detected 2
The detection of the
TIPCAP failed.
The activity is aborted.
- Remove the inlet flaps
and the inlet gasket.
Inspect the condition of the
inlet probe and gasket.
Look for obstructions that
may hinder probe
movement. Replace as
necessary. Install all parts.
- Perform the Inlet Probe
Troubleshooting
Procedure.
- Restart the analyzer.
- Contact Radiometer
Service Representative.
Removal condition:
Successful inlet calibration.
Continued on next page
11-62