11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
937 Cleaning
Solution -
Crea will
expire soon
The additive used in
the Cleaning Solution
to measure the High
Crea Check will expire
soon.
This is a warning - the
solution lifetime is 14
days.
Replace the solution before
the lifetime is actually
exceeded.
938 Cal1
Solution -
Crea is
expired
The additive used in
the Calibration
Solution 1 to calibrate
Crea is expired.
The solution is more
than 14 days old.
It is no longer possible
to calibrate Crea.
Replace the Calibration
Solution 1 bottle -
remember to add the
additive.
Removal condition:
Fresh bottle installed.
939 Cal2
Solution -
Crea is
expired
The additive used in
the Calibration
Solution 2 to calibrate
Crea is expired.
The solution is more
than 14 days old.
It is no longer possible
to calibrate Crea.
Replace the Calibration
Solution 2 bottle -
remember to add the
additive.
Removal condition:
Fresh bottle installed.
940 Cleaning
Solution -
Crea is
expired
The additive used in
the Cleaning Solution
to measure the High
Crea Check is expired.
The solution is more
than 14 days old.
It is no longer possible
to perform the High
Crea Check.
Replace the Cleaning
Solution bottle - remember
to add the additive.
Removal condition:
Fresh bottle installed.
Continued on next page
11-64