ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
941 Cal expired
(Crea)
942 Cal expired
(Creatine)
Too long time passed
since the last successful
calibration of the
parameter. Parameter
measurement values
are reported as ".....".
Perform a 2-point
calibration.
Removal condition:
Successful 2-point
calibration.
944 Met II
module not
expected
Software configuration
does not match
analyzer type.
Contact Radiometer
Service Representative.
Removal condition:
Matching software and
hardware configuration.
945 Met II
module not
active
The Met II module is
not responding due to
internal communication
problems, or software
configuration does not
match the analyzer
type.
- Shut down the analyzer
using the Temporary
Shutdown function, then
restart it.
- Contact Radiometer
Service Representative.
Removal condition:
Met II module ready, or
software configured
without Met II module
support.
946 Conditioning
error in Met
II electrode
module
Rinse Solution was not
detected in Met II
electrode module. The
Crea results will be
reported as "....." in the
next measurement.
- Check Rinse Solution for
adequate volume, proper
installation or obstructions
in bottle opening.
- Replace as necessary.
- Perform the Fluid
Transport Troubleshooting
procedure.
Removal condition:
Succesful Rinse.
Continued on next page
11-65