11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
951 Creatine is
below the
reportable
range
Creatine is below the
reportable range.
See chapter 14:
Specifications in the
Operator's Manual.
- Check for and remedy
other errors related to the
result, system messages or
calibration status.
- Perform QC. If the QC
result is accepted, the
blood sample may be
suspected.
- Perform measurement on
new blood sample.
- Perform the Electrode
Troubleshooting procedure.
952 Creatine is
above the
reportable
range
Creatine is above the
reportable range.
See chapter 14:
Specifications in the
Operator's Manual.
- Check for and remedy
other errors related to the
result, system messages or
calibration status.
- Perform QC. If the QC
result is accepted, the
blood sample may be
suspected.
- Perform measurement on
new blood sample.
- Perform the Electrode
Troubleshooting procedure.
953 Measured
QC value
above the
control
range
(Crea A)
Creatine value is above
the control range.
Error on the Creatine
measurement is
possible.
- Verify procedure and
repeat measurement.
- Refer to Quality Control
Systems Reference
Manual.
954 Measured
QC value
below the
control
range
(Crea A)
Creatine value is below
the control range.
Error on the Creatine
measurement is
possible.
- Verify procedure and
repeat measurement.
- Refer to the Quality
Control Systems Reference
Manual.
Continued on next page
11-66