ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
955 Measured
QC value
lower than
statistical
range
(Crea A)
The parameter value is
below the lower limit
of the user-defined
statistical range.
Measurement is not
included in the
statistics.
- Verify procedure and
repeat the measurement.
- Refer to the Quality
Control Systems Reference
Manual for details on the
evaluation of the results.
956 Measured
QC value
higher than
statistical
range
(Crea A)
The parameter value is
above the upper limit
of the user-defined
statistical range.
Measurement is not
included in the
statistics.
- Verify procedure and
repeat the measurement.
- Refer to the Quality
Control Systems Reference
Manual for details on the
evaluation of the results.
958 Solution
lifetime
exceeded
A solution is more than
14 days old and cannot
be used for Crea
calibration or High
Crea Check.
Calibration Solutions:
- Crea is no longer
reported during
measurements and is
marked red on the
parameter bar.
Cleaning Solution:
- It is no longer
possible to perform a
High Crea Check.
Replace required solution
bottle - remember to add
the additive.
Removal condition:
Fresh bottle installed.
Continued on next page
11-67