ABL800 FLEX Operator's Manual 11. Troubleshooting
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
961 Total
concentration
outside
measuring
range
The total concentration
of Crea + Creatine is
above the measuring
range and the analyzer
is not able to measure
Crea within the
specifications.
See chapter 14:
Specifications in the
Operator's Manual.
- Check for and remedy
other errors related to the
result, system messages or
calibration status.
- Perform a High Crea
Check. If the High Crea
Check result is accepted,
the blood sample may be
suspected.
- Perform measurement on
new blood sample.
- Perform the Electrode
Troubleshooting procedure.
962 Ambient
temperature
not defined
Ambient temperature
has not been defined by
the user in
Environment Setup.
Until the ambient
temperature has been
defined, a default value
of 25 °C is used during
High Crea Check and
the internal Crea
calibration calculations.
Please maintain an
ambient temperature
between 18 °C and 29
°C
Enter ambient temperature
in Environment Setup - see
Operator's Manual chapter
3.
Removal condition:
Ambient temperature
entered.
963, 964 Leak
currents
detected
Leak currents are
detected during
calibration and may
distort the measuring
results. The affected
parameters are marked
with "?".
Contact Radiometer service
representative.
965 Cleaning
Additive is
expired
The Cleaning additive
is too old and the
Cleaning Solution
cannot be used
anymore.
Replace Cleaning Solution
bottle – remember to add
the additive.
Continued on next page
11-69