11. Troubleshooting ABL800 FLEX Operator's Manual
Analyzer messages, Continued
Messages at
manager level
(continued)
No. Message Interpretation Operator action
966 Cleaning –
Additive
will expire
soon
The Cleaning additive
will expire soon.
Replace the solution before
the lifetime is actually
exceeded.
967 Creatine
high.
Unable to
perform
Crea
measure-
ment
The Creatine
concentration is too
high compared to the
Crea concentration and
the analyzer is not able
to measure Crea within
the specifications.
See chapter 14:
Specifications in the
Operator's Manual
- Check for and remedy
other errors related to the
result, system messages or
calibration status.
- Perform a High Crea
Check. If the High Crea
Check result is accepted,
the blood sample may be
suspected.
- Perform measurement on
a new blood sample.
- Perform the Electrode
Troubleshooting procedure
969 FLEXQ
module not
expected
Software configuration
does not match
analyzer type.
Contact Radiometer
Service Representative.
Removal condition:
Matching software and
hardware configuration.
970 Rilibäk
violation:
Value above
upper limit
971 Rilibäk
violation:
Value belov
lower limit
Continued on next page
11-70

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADIOMETER ABL800 FLEX and is the answer not in the manual?

RADIOMETER ABL800 FLEX Specifications

General IconGeneral
BrandRADIOMETER
ModelABL800 FLEX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals