ABL800 FLEX Operator's Manual 2. What Is what
Analyzer status, Continued
Quality Control
status
The Quality Control status provides the following information:
•
status of the last measurement on each control solution type
•
messages referring to quality control measurements.
For the ABL8x7 FLEX analyzers, please note that Slot 11 will be automatically
installed when you scan the barcode for the Cleaning Met II Solution during
replacement.
Element Function
Solution Lists the quality control slot and its solution type along
with a status symbol.
Cleaning Met II Solution (for ABL8x7 FLEX analyzers
only) is used for quality control of high creatinine levels.
OK
The last measurement was accepted.
?
One or more of the following occurred:
• error in the last calibration.
• analyzer error during last QC measurement.
• a parameter measurement is outside the defined
ranges or a Westgard Rule has been violated.
The next measurement is overdue, and the
previous measurement, if any, was accepted.
?
The last quality control measurement had errors
present, and the next measurement is overdue.
Lot The solution lot number for the slot.
Last Time Lists the time the last measurement was performed.
Next Time Lists the next scheduled time to perform the measurement
on the slot. See Quality Control Schedule, chapter 3.
Continued on next page
2-31