ABL800 FLEX Operator's Manual 3. Installation and setup
Analyzer security, Continued
General
Security
(continued)
To define, how the user should log on, use the up/down arrow buttons in the
"Authenticate" box to select the desired logon option. The following options are
available:
• UserId/Password as primary
This option allows you to enter or scan a User name and password in the Logon
screen. By pressing the Log On BC button a Logon-barcode can be scanned.
• UserId/Password only
This option allows you to enter or scan the user name and password in the
Logon screen.
• Logon-barcode as primary
This option allows you to enter or scan a Logon-barcode in the Logon screen.
By pressing the Extented Log On button a user name and password can be
scanned.
• Logon-barcode only
This option allows you to enter or scan a Logon-barcode in the Logon screen.
To allow an anonymous use of the analyzer, i.e. use without logon, use the
up/down arrow buttons to select "Yes" in the "Anonymous use" box (default) and
select the desired access profile of the anonymous operator, in the "Access profile
for anonymous operator" box. The "Access profile for anonymous operator" box
only appears, when "Yes" is chosen in the "Anonymous use" box.
See Access Profiles further in this chapter for information on how to define the
access profiles.
To set the time interval to elapse, before an operator is automatically logged off,
press the Logoff Time button. Select the logoff time in minutes (from 0 to 60) and
seconds (from 0 to 50 in 10-second intervals). The default logoff time is three
minutes. Press Back to return to the General Security screen.
Continued on next page
3-7