3. Installation and setup ABL800 FLEX Operator's Manual
Analysis setup, Continued
Selecting a
measuring
program
To select a measuring program, do the following:
Step Action
1.
Blood, pleura and expired air samples
: Press the desired mode button,
enable it and select the desired measuring program with the arrow
buttons.
Other fluids:
Activate the desired mode button and enable the Other
fluids check button.
2.
Press Parameters to select a parameter profile for a measuring
program − see Selecting parameter profile below.
3.
Edit, if desired, the left part of the button name for the mode by
pressing Edit Name.
4.
Enter corrections for the Other fluids mode by pressing Corrections −
see Entering corrections for Other Fluids Modes further in this
section.
5.
Select the report layout for a measuring program by pressing Layout −
see Selecting a report layout further in this section.
NOTICES:
•
You can use one of the syringe measuring modes for test
tubes.
•
Use Capillary − 35 μL OXI for bilirubin analysis (ctHb = 0)
on plasma. Deselect all other parameters for this mode in the
Parameter Profile. Alternatively, you can use the Other Fluids
mode for the bilirubin measurement on plasma (this is not
relevant for the ABL8x7 FLEX analyzers).
•
In the Syringe Modes Setup and Capillary Modes Setup
screens you can define up to 12 measuring programs. To get
access to the last 6 mode buttons press Next Page.
Continued on next page
3-14