ABL800 FLEX Operator's Manual 3. Installation and setup
Analysis setup, Continued
Entering
corrections for
Other Fluids
mode
The Other Fluids mode allows measurement on samples other than heparinized
whole human blood and, therefore, requires entering corrections (offset and slope).
Step Action
1.
Select the Other Fluids mode and press the Corrections button on one
of the Modes Setup screens (Capillary or Syringe).
2.
Highlight the parameter with the arrow buttons and press Edit.
3.
Key in the slope and offset values. Confirm each entry with Enter.
Please note that only values within the range programmed into the
software will be accepted.
Press Back to confirm the entries and return to the previous screen.
4.
5.
To delete a correction, highlight it and then press the
Remove
button.
Continued on next page
3-17