3. Installation and setup ABL800 FLEX Operator's Manual
Analysis setup, Continued
Selecting HbF
correction
To give the user the possibility of selecting or deselecting HbF correction during a
measurement, do the following:
Step Action
1.
Select HbF correction for all levels or for HbF levels higher than 20 %
in Miscellaneous Setup (Setup General Setup Miscellaneous
Setup) – see the description further in this chapter.
2.
Return to the Analysis Setup, choose the desired syringe or capillary
program and press the Parameters button to enter the Parameter
Profile screen.
3.
Activate the check button in the "Use HbF correction" box (if
"Disabled for all levels" has been selected in Miscellaneous Setup, the
text box and the check button in it will not be visible on this screen).
4.
Press the dynamic parameters check button in the “Use dynamic
parameters” box in order to be able to exclude the HbF correction and
other parameters during a sample measurement.
HbF correction
was disabled in
Miscellaneous
Setup